Abstract

The National Radiological Protection Board (NRPB) has recommended the introduction of dose-width product (DWP) for the measurement of patient dose in panoramic dental radiology and has proposed a reference level of 65 mGy mm for adult exposures. This paper describes a method for measuring DWP and dose-area product (DAP) using thermoluminescent dosemeters (TLDs). The technique was used on 16 sets with a range of exposure settings. The mean value of DWP was 14% higher than the mean value reported from a survey by the NRPB. This difference is most likely to be caused by systematic variations due to measurement method. The average DAP for a standard adult examination was shown to be 11.3 cGy cm2. Data are presented so that the DAP can be derived from the exposure factors (tube current and operating potential) and beam area. Based on published data for effective dose, it is estimated that the DAP to effective dose conversion factor is approximately 0.06 mSv(Gy cm2)-1. The average DAP value (11.3 cGy cm2) can be compared with the average value for intraoral radiography (9.3 cGy cm2) based on the NRPB survey of entrance surface doses assuming 6 cm circular collimation.