Abstract

Common left coronary blood flow was measured simultaneously by a recording rotameter and by the Rb 86 chloride clearance method in anesthetized, open chest dogs, and compared over a wide range of flow rates and under conditions of both normal and abnormal myocardial oxygenation and function. Although the Rb 86 clearance method reflected the same directional changes as metered flow, a systematic underestimate was found which ranged from 27% at ischemic flow rates to 54% at hyperemic rates. Myocardial extraction studies supported the view that coronary flow rate and the resultant time that the isotope was in the capillary bed was the major determinant of the underestimate by the Rb 86 method even under conditions of myocardial hypoxia. It was concluded that the Rb 86 method gave an accurate estimate of directional change in coronary blood flow but that the magnitude of the underestimate of actual flow limits its use as a quantitative index.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.