Abstract

Background: Currently, the DRL quantity in mammography are evaluated in terms of mean glandular dose (MGD). Since the MGD cannot be measured directly, it can be obtained by calculation using the equation (D=K*g*c*s). In previous studies, the conversion factor g was calculated by Monte Carlo simulation and is not reported from actual measurements. In this study, we focused on the g-factor, which is a conversion factor to the MGD at 50% glandularity, and attempted to measure it using a nanoDot dosimeter to see if it can be used in mammography. Methods: The nanoDot dosimeters were inserted in a PMMA phantom at depths ranging from 0 cm to 6 cm in 1 cm increments, and measurements were made in three HVLs of 0.3 mmAl, 0.35 mmAl, and 0.4 mmAl HVL. The g-factor was calculated from the nanoDot dosimeter values using a conversion equation. Results and Discussion: The measured g-factors for all the HVLs were in close agreement with those of Dance et al. The values of the previous studies did not include the backscatter factor, which may have underestimated the MGD. The difference was smaller for the 0.4 mm Al. Compared to the other HVLs, the 0.4 mm Al was measured without a compression plate, which may have been influenced by the presence or absence of a compression plate. Conclusion: The nanoDot dosimeters were used to calculate g-factors. The results agreed with those of previous studies within uncertainty. This indicates that nanoDot dosimeters can be used in the mammography field.