Abstract

Total dissolved inorganic carbon (SigmaCO(2)) and aqueous carbon dioxide (H(2)CO(3) (*)) in nutrient solutions may be measured by the injection of small gas or liquid samples (1 microliter to 8 milliliters) into a gas stripping column connected in-line with an infrared gas analyzer. The measurement of SigmaCO(2) in solution requires sample acidification, while H(2)CO(3) (*) and gaseous CO(2) are measured without the addition of lactic acid. The standard curve for SigmaCO(2) was linear up to 300 nanomoles CO(2). Maximum sensitivity was approximately 300 picomoles. Measurements of H(2)CO(3) (*) were independent of pH. Consequently, SigmaCO(2) and H(2)CO(3) (*) could be used to calculate the pH, HCO(3) (-), and CO(3) (2-) values of nutrient solutions. Injection and complete analyses required from 0.8 to 2 minutes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.