Abstract

The adsorption of carbon dioxide (CO 2 ) on solid sorbents or any gas adsorption in general can be experimentally determined using various techniques. The present study demonstrates a dynamic column breakthrough method for the measurement of CO 2 adsorption using a cost-effective purpose-built single-column fixed-bed adsorption unit. The CO 2 adsorption capacity is estimated using a graphical method that does not require knowledge of the variation of molar CO 2 flow rate at the column exit. The experimental method is validated using a commercial zeolite 13X adsorbent.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.