Abstract

X-ray doses in the range of centi-Gray can be detected and accurately measured on a relative scale for low energy X-rays using a high sensitivity radiochromic film stack. Gafchromic XR type T radiochromic film has been investigated to analyze its ability to measure relative doses below 50 cGy for 100-kVp X-rays using a layered technique. Investigations showed that a 4-layer film stack could produce reproducible results at 50-cGy relative applied dose to within 3% and to within 12% for 5-cGy applied dose. The combination of single sheet higher film sensitivity to low energy X-rays along with a layered film dosimetry technique provides the ability to adequately measure low relative absorbed doses less than 50 cGy if required.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.