Abstract

X-ray doses in the range of centi-Gray can be detected and accurately measured on a relative scale for low energy X-rays using a high sensitivity radiochromic film stack. Gafchromic XR type T radiochromic film has been investigated to analyze its ability to measure relative doses below 50 cGy for 100-kVp X-rays using a layered technique. Investigations showed that a 4-layer film stack could produce reproducible results at 50-cGy relative applied dose to within 3% and to within 12% for 5-cGy applied dose. The combination of single sheet higher film sensitivity to low energy X-rays along with a layered film dosimetry technique provides the ability to adequately measure low relative absorbed doses less than 50 cGy if required.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.