Abstract
The main goal of this work was to describe the analytical characteristics of an enzymatic photometric test for carnitine determination and its automation using an Olympus analyzer. We used a test from Roche intended for manual processing and tried to apply it for use on an Olympus AU 400 analyzer. The analytical parameters of our modified technique were determined using external quality controls and kit controls, and by measurements in venous blood samples from 85 chronically hemodialyzed patients (before and after hemodialysis) and from 68 healthy blood donors serving as controls. A reference value for free carnitine was estimated parametrically as 40.1+/-17.8 micromol/L. The mean bias for eight control measurements was 5.1%. Sensitivity was calculated as the limit of quantification at 2.6 micromol/L. The intra-assay coefficient of variation was 2.4%. The inter-assay coefficient of variation was 8.3%. Analytical recovery was 101.8%, 99.5% and 95.4%. The main advantages of our automated method in comparison to the original manual method are the smaller amounts of samples, reagents and diluents required and the shorter analysis time. As hemodialysis patients often suffer from carnitine deficiency, we conclude that its determination may be helpful for diagnostic verification.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.