Our aim was to investigate the bone thickness at the site of titanium miniplates inserted to retain nasal prostheses. We studied 13 patients who had had titanium miniplates inserted for retention of nasal prostheses with a total of 60 titanium bone screws. A trajectory along each bone screw was segmented in fused computed tomographic (CT) data. Bone thickness was measured along this trajectory on the preoperative CT. The median bone thickness at the positions of the screws implanted on the frontal process of the maxillary bone was 1.4 (range 0.2-6.9) mm (mean 1.8). The median (range) values for men and women were 1.4 (0.2-6.9) mm and 1.3 (0.2-3.3) mm, respectively. The thickest bone was at the cranial part of the frontal process of the maxilla with a median of 2.0 (range 0.3-4.1) mm. However, differences in sex and position were not significant. None of the implant miniplates lost osseointegration. Despite the low bone stock at the lateral aspect of the pyriform aperture, survival of implanted titanium miniplates was 100% in this study group.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.