Abstract

Capillary blood sampling as a means of monitoring blood cyclosporine levels has replaced venipuncture in some medical centers. As the validity of capillary venipuncture for analysis of cyclosporine has not been documented, we sought to validate the capillary blood collection technique by comparing it with serum samples collected simultaneously by venous phlebotomy. Forty paired capillary- and venous samples were collected from 36 cardiac transplantation patients and analyzed, using a polyclonal immunoassay. The values obtained were compared using regression correlation. The correlation coefficient for all 40 samples was 0.859. However, we discovered that the first 7 capillary specimens were processed incorrectly. The correlation coefficient for the other 33 samples was 0.995 (99% confidence interval 0.987-0.998). The excellent correlation between serum samples obtained from capillary sampling and from venous sampling, together with the ease of obtaining capillary blood specimens, make "fingerstick" sampling the method of choice for monitoring cyclosporine levels in infants and children.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.