Abstract

The purpose of this study is to establish an easy and accurate method for the determination of bisphenol A (BPA) in the human serum. The samples were applied to the C 18 solid phase extraction (SPE) column for clean up of samples. The BPA is conjugated with tetrabutylammonium hydrogen sulfate as the counter ion in alkali solution. The ion paired BPA is moves from the aqueous phase to the organic phase as an ion paired extraction. BPA extracted from human serum were derivatized with pentafluorobenzyl bromide (PFBBr). The derivative was analyzed by gas chromatography (GC)/mass spectrometry (MS) using negative chemical ionization (NCI). The instrumental detection limit of BPA was 5 pg/ml (10 fg). The instrumental response between 0.01 and 100 pg/ml of BPA standards was linear ( r 2=0.998). The recovery of BPA spiked into human serum was 101.0±4.63 (1 pg/ml) and 100.9±3.75 (10 pg/ml), respectively. The concentration of BPA in the human serum from 20 individuals was 0.54 pg/ml.