Abstract

The basic characteristics of three low-cost radiation survey meters using multi-pixel photon counters (MPPC), the C12137, T-GMK2-S, and iMetry, were measured. The linearity of the dose rate was confirmed over the full range of each instrument. All the survey meters could obtain gamma-ray energy spectra, with an accuracy within ± 30% of the theoretical value. These survey meters are therefore applicable for radiation management.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.