Abstract

The Magnetic Recoil neutron Spectrometer (MRS) at the OMEGA laser facility has been routinely used to measure deuterium-tritium (DT) yield and areal density in cryogenically layered implosions since 2008. Recently, operation of the OMEGA MRS in higher-resolution mode with a new smaller, thinner (4 cm2, 57 μm thick) CD2 conversion foil has also enabled inference of the apparent DT ion temperature (T ion) from MRS data. MRS-inferred T ion compares well with T ion as measured using neutron time-of-flight spectrometers, which is important as it demonstrates good understanding of the very different systematics associated with the two independent measurements. The MRS resolution in this configuration, ΔE MRS = 0.91 MeV FWHM, is still higher than that required for a high-precision T ion measurement. We show how fielding a smaller foil closer to the target chamber center and redesigning the MRS detector array could bring the resolution to ΔE MRS = 0.45 MeV, reducing the systematic T ion uncertainty by more than a factor of 4.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call