Abstract

Certain odor control regulations specify use of the Scentometer for ambient odor measurement. This evaluation is usually performed by a single individual who is surrounded by the odorous environment to be measured. A method is desired where an ambient odor sample can be evaluated by an adequate size panel in an odor-free atmosphere. A dynamic forced-choice triangle olfactometer was designed and constructed to measure ambient odors. Teflon bags of 18 liter capacity collect a sample within 2-3 minutes which includes pre-flushing the bag. The sample is evaluated by a dynamic olfactometer equipped with 5 dilution levels (81×, 27×, 9×, 3× and undiluted sample). Three sniffing ports are provided at each dilution level to present dynamically one diluted odor stimulus and two odor-free air blanks. Each panelist is required to indicate which port contains the odor. Evaluation of one sample is routinely completed by a panel of 9 within less than 15 minutes. The odor threshold value (ED50) for the panel is calculated by use of a simple table derived statistically. No significant loss of odor was observed in sampling and in storage of rendering odors up to 48 hours. Bags were reusable after flushing with odor-free air. Reproducibility of log ED50 values by the same panel was within a σ = 0.10 log10. Agreement in evaluating duplicate field samples by two different panels was within the same limits. Under controlled laboratory conditions, a Scentometer reading of D/T = 2 was equivalent to an ED50 = 4.8; and D/T = 7 was equal to ED50 = 9.5.