Abstract

The aim of this study was to compare measurements of angle of lateral opening (ALO) and version determined using a radioopaque cup position assessment device imaged with fluoroscopy to measurements obtained by CT and direct measurement in a cadaveric model. Our null hypothesis was that there would not be any difference in the angles measured by the techniques. Six cadavers were implanted with BFX acetabular components. The CPAD was placed and images were obtained with fluoroscopy. Measurements were obtained from the radiopaque marker bars on the CPAD device, and version and ALO were calculated. The ALO and version were determined by CT and DM. Comparisons were made using a two-way analysis of variance and a generalized linear model procedure analysis. There were no significant differences between the measurements for ALO (p = 0.275) or version (p = 0.226). Correlation between methods was 0.948 and 0.951 for ALO and version, respectively. The mean difference (standard deviation [SD], and 95% confidence interval [CI]) for ALO were: CT versus CPAD 1.85 degrees (± 2.32 degrees [-2.99-3.31]), CT versus DM 1.96 degrees (± 1.99 degrees [-2.2-4.27]), CPAD versus DM1.74 degrees (±2.21 degrees [-1.13 and 5.24]). The mean difference (SD [CI]) for version was CT versus CPAD 2.86 degrees (±1.56 degrees [ -2.63-1.69]), CT versus DM 1.10 degrees (±1.42 degrees [-1.57-2.09]), CPAD versus DM 1.07 degrees (±0.76 degrees [0.13-2.09]). The results demonstrate that intraoperative imaging in cadaveric specimens with the CPAD is an accurate method to determine ALO and version of the acetabular component.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call