Abstract
RationaleThe doubly substituted isotopologues (e.g., 18O18O, 17O18O) in atmospheric O2 are potential tracers for ozone photochemistry and atmospheric temperatures. Their low abundances and isobaric interference are the major analytical challenges. The 253 Ultra high‐resolution stable isotope ratio mass spectrometer is suitable for resolving isobaric interferences.MethodsO2 from air is purified using gas chromatography on a packed column filled with molecular sieve 5 Å and cooled to −78°C. The δ17O, δ18O, Δ17O, Δ35 and Δ36 values are measured on the extracted O2 with the 253 Ultra at medium mass resolution (M/ΔM ~10000) using Faraday detectors for the singly substituted isotopologues and ion counters for the doubly substituted isotopologues.ResultsInterferences from isobars, mainly 35Cl for 17O18O and H35Cl and 36Ar for 18O18O, are sufficiently resolved to enable high‐precision determination of Δ35 and Δ36. The Δ35 and Δ36 values of O2 after photochemical isotope equilibration at −63°C and heating to 850°C agree with the theoretical prediction. The stratospheric Δ35 and Δ36 values are close to isotopic equilibrium at the ambient temperatures. However, the values for tropospheric O2 differ from those expected at equilibrium.ConclusionsThe 253 Ultra allows interference‐free clumped isotope measurements of O2 at medium mass resolution. The Δ35 and Δ36 signatures in atmospheric O2 are mainly governed by O3 photochemistry, temperature and atmospheric transport. Tropospheric O2 is isotopically well mixed and retains a significant stratospheric signature.
Highlights
We show that traces of 36Ar and other interferences can be well separated from 17O18O and 18O18O at medium mass resolution (~10000)
We report Δ35 and Δ36 values measured in stratospheric and tropospheric air O2 samples
Isotopic measurements including the clumped isotopes (Δ35 and Δ36) were carried out with the 253 Ultra isotope ratio mass spectrometer at Utrecht University, which is one of the first instruments produced by Thermo Fisher Scientific on the basis of the prototype described in Eiler et al.[16]
Summary
O2 from air is purified using gas chromatography on a packed column filled with molecular sieve 5 Å and cooled to −78°C. The δ17O, δ18O, Δ17O, Δ35 and Δ36 values are measured on the extracted O2 with the 253 Ultra at medium mass resolution (M/ΔM ~10000) using Faraday detectors for the singly substituted isotopologues and ion counters for the doubly substituted isotopologues. The Δ35 and Δ36 values of O2 after photochemical isotope equilibration at −63°C and heating to 850°C agree with the theoretical prediction. The stratospheric Δ35 and Δ36 values are close to isotopic equilibrium at the ambient temperatures. The values for tropospheric O2 differ from those expected at equilibrium. Conclusions: The 253 Ultra allows interference‐free clumped isotope measurements of O2 at medium mass resolution. The Δ35 and Δ36 signatures in atmospheric O2 are mainly governed by O3 photochemistry, temperature and atmospheric transport. Tropospheric O2 is isotopically well mixed and retains a significant stratospheric signature
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.