Abstract

We measured tissue temperature changes during ex vivo microwave ablation (MWA) procedures for bovine liver tissue. Tissue temperature increased rapidly at the beginning of the MW power application. It came to a plateau at 100 degrees C to 104 degrees C before it increased again. We split the changes of tissue temperature versus time into four phases. This suggests that tissue temperature changes may be directly related to tissue water related phenomena during MWA, including evaporation, diffusion, condensation and tissue water composition. An additional analysis indicated the lesion boundary at approximately 50 degres C to 60 degrees C temperature. We also measured the water content of ablated tissue lesions and examined the relationship of tissue water content and tissue temperature by mapping temperature to remaining tissue water after ablation. The results demonstrate significant tissue water content changes and lead to a better understanding of tissue water movement.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.