Abstract

The purpose of this paper was to estimate mean glandular dose levels (DG) in six digital mammography systems in Santiago, Chile, and to propose preliminary reference levels to execute mammography in Chile. The study was carried out assessing two direct digital systems and four computer-based radiography (CR) systems. Estimates of DG were calculated for different thicknesses of polymethyl methacrylate according to the quality control protocol in digital mammography of the Spanish Society of Medical Physics and NHSBSP Equipment Report 0604 Version 3. DG values ranged between 0.64 and 7.26 mGy for a range of 20- to 70-mm thickness, respectively. Thirty-six per cent of DG was higher than the acceptable dose level and 100 % of DG was higher than the desirable level. It is therefore necessary to optimise doses. The initial proposal to establish dose reference levels for DG would range between 0.90 and 6.40 mGy for a thickness range of 20 to 70 mm.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.