Abstract

This work presents the preliminary results obtained with a high-resistivity magnetic Czochralski (MCz) silicon diode processed at the Helsinki Institute of Physics as a high-dose gamma dosimeter in radiation processing. The irradiation was performed using a 60Co source (Gammacell 220, MDS Nordion) within total doses from 100 Gy up to 3 kGy at a dose rate of 3 kGy/h. In this interval, the dosimetric response of the diode is linear with a correlation coefficient ( r 2 ) higher than 0.993. However, without any irradiation procedure, the device showed a small sensitivity dependence on the accumulated dose. For total dose of 3 kGy, the observed decrease was about 2%. To clarify the origin of this possible radiation damage effect, some studies are under way.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.