Abstract

We report here an optimisation of the demountable direct injection high efficiency nebuliser (d‐DIHEN) for isotopic measurements with a Neptune (ThermoFisher Scientific, Bremen, Germany) multi‐collector inductively coupled plasma‐mass spectrometer (MC‐ICP‐MS) and describe a method for boron isotopic ratio determination. With direct injection nebulisation 100% of the analyte was introduced into the ICP‐MS plasma and wash times were drastically reduced for elements such as boron and thorium. Compared to the classical stable introduction system (SIS: double Scott/cyclonic spray chamber), sensitivity for boron was 2–5 times higher with d‐DIHEN and wash times up to ten times shorter. Repeatability of 11B/10B sample‐calibrator bracketing measurements reached 0.25‰ (2s) for seawater and coral samples. Method accuracy and reproducibility were tested on mixed reference solutions having δ11B values in the ranges −90 to +40‰ and −2 to +2.5‰, demonstrating our ability to distinguish δ11B values with differences of only 0.25‰. The international seawater reference material NRCC NASS‐5 (National Research Council, Ottawa, Canada), analysed in different sessions over a 10‐month period, yielded an average δ11B value of +39.89 ± 0.25‰, in the upper range of previously published seawater values. A comparison between δ11B determined by d‐DIHEN MC‐ICP‐MS and positive‐TIMS (P‐TIMS) for four modern corals showed an excellent agreement (with bias of less than 0.4‰).