Abstract

Background. Three commonly used sutures were tested in a pig model of bile duct anastomosis to assess their relative contributions to inflammation and scarring. Methods. Thirty pigs were randomised to bile duct division and anastomosis with either polyglyconate (Maxon), polyglactin 910 (Vicryl) or polypropylene (Prolene). Half the animals were sacrificed at two weeks and the remainder at 23 weeks. Anastomoses were assessed by cholangiography, scanning electron microscopy and light microscopy. Results. There was less short term histological reaction with the two monofilament materials, Prolene and Maxon, compared to the braided suture Vicryl. Maxon was associated with less long term inflammation than Prolene, was found to handle better, and has an advantage over Prolene by being absorbable. Conclusion. Maxon is an optimal suture for bile duct anastomoses. Its long term absorption characteristics make it suitable for situations where bile duct healing may be delayed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.