Abstract

Parametric imaging using Patlak graphical method has been widely used to analyze dynamic PET data. The conventional way to generate Patlak parametric image is to reconstruct dynamic images first and then perform Patlak graphical analysis on the time activity curves pixel-by-pixel. In this paper we present a Bayesian method for reconstructing Patlak parametric images directly from raw sinogram data by combining the Patlak plot model with image reconstruction. A preconditioned conjugate gradient algorithm is used to find the maximum a posteriori solution. We conduct computer simulations to validate the proposed method. The comparison with conventional indirect approaches shows that the proposed method results in more accurate estimate of the parametric image

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.