Abstract

We examined the ratio between actual maxillary distraction and the distraction of the rigid external distraction device (Rigid external distraction (RED) system) used for maxillary distraction in patients with a cleft deformity. Twelve patients were examined. The amount of maxillary advancement was measured on lateral cephalograms and divided by the activation amount on the RED system. The value obtained was represented as the distraction ratio of the maxilla to the system. The mean ratio in 10 patients with complete cleft lip, palate, and alveolus (complete cleft) was 0.24. However, the ratios in two patients with cleft lip and alveolus or soft cleft palate (incomplete cleft) were considerably higher than the mean ratios in patients with complete cleft. When the maxilla is distracted in patients with complete cleft using the RED system, the amount of activation on the system needs to be about four times the amount of planned maxillary distraction. However, the distraction ratio may be affected by the type of cleft.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.