Abstract

Brominated analogues (BMXs) of the strong drinking water mutagen MX (3-chloro-4-(dichloromethyl)-5-hydroxy-2(5 H)-furanone) were found to be subject to strong matrix induced chromatographic response enhancement effects. We evaluated different ways to reduce errors in quantification including comparison of gas chromatographic inlet systems, improved clean up of sample extracts, and preparation of calibration standards in the sample matrix. The best quantitative accuracy and long term performance were achieved when the calibration standards were prepared in sample matrix, samples were cleaned up with C 18-resin in conjunction with solid phase extraction (SPE) with Oasis HLB cartridges, and gas chromatography with PTV splitless injection was used. This method enables the determination of MX and BMXs from 500 ml water sample with quantitation limits of 1 ng/l or less.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.