Abstract

Stenosis-induced thrombosis and abandonment of the hemodialysis synthetic graft is an important cause of morbidity and mortality. The graft vascular circuit is a unique low-resistance shunt that has not yet been systematically evaluated. In this study, we developed a mathematical model of this circuit. Pressure losses (deltaPs) were measured in an in vitro experimental apparatus and compared with losses predicted by equations from the engineering literature. We considered the inflow artery, arterial and venous anastomoses, graft, stenosis, and outflow vein. We found significant differences between equations and experimental results, and attributed these differences to the transitional nature of the flow. Adjustment of the equations led to good agreement with experimental data. The resulting mathematical model predicts relations between stenosis, blood flow, intragraft pressure, and important clinical variables such as mean arterial blood pressure and hematocrit. Application of the model should improve understanding of the hemodynamics of the stenotic graft vascular circuit.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.