Abstract
The full-energy peak efficiency of high-purity germanium well-type detector is extremely important to calculate the absolute activities of natural and artificial radionuclides for samples with low radioactivity. In this work, the efficiency transfer method in an integral form is proposed to calculate the full-energy peak efficiency and to correct the coincidence summing effect for a high-purity germanium well-type detector. This technique is based on the calculation of the ratio of the effective solid angles subtended by the well-type detector with cylindrical sources measured inside detector cavity and an axial point source measured out the detector cavity including the attenuation of the photon by the absorber system. This technique can be easily applied in establishing the efficiency calibration curves of well-type detectors. The calculated values of the efficiency are in good agreement with the experimental calibration data obtained with a mixed γ-ray standard source containing 60Co and 88Y.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.