Abstract

BackgroundThe wide range of fresh gas flow - vaporizer setting (FGF - FD) combinations used by different anesthesiologists during the wash-in period of inhaled anesthetics indicates that the selection of FGF and FD is based on habit and personal experience. An empirical model could rationalize FGF - FD selection during wash-in.MethodsDuring model derivation, 50 ASA PS I-II patients received desflurane in O2 with an ADU® anesthesia machine with a random combination of a fixed FGF - FD setting. The resulting course of the end-expired desflurane concentration (FA) was modeled with Excel Solver, with patient age, height, and weight as covariates; NONMEM was used to check for parsimony. The resulting equation was solved for FD, and prospectively tested by having the formula calculate FD to be used by the anesthesiologist after randomly selecting a FGF, a target FA (FAt), and a specified time interval (1 - 5 min) after turning on the vaporizer after which FAt had to be reached. The following targets were tested: desflurane FAt 3.5% after 3.5 min (n = 40), 5% after 5 min (n = 37), and 6% after 4.5 min (n = 37).ResultsSolving the equation derived during model development for FD yields FD=-(e(-FGF*-0.23+FGF*0.24)*(e(FGF*-0.23)*FAt*Ht*0.1-e(FGF*-0.23)*FGF*2.55+40.46-e(FGF*-0.23)*40.46+e(FGF*-0.23+Time/-4.08)*40.46-e(Time/-4.08)*40.46))/((-1+e(FGF*0.24))*(-1+e(Time/-4.08))*39.29). Only height (Ht) could be withheld as a significant covariate. Median performance error and median absolute performance error were -2.9 and 7.0% in the 3.5% after 3.5 min group, -3.4 and 11.4% in the 5% after 5 min group, and -16.2 and 16.2% in the 6% after 4.5 min groups, respectively.ConclusionsAn empirical model can be used to predict the FGF - FD combinations that attain a target end-expired anesthetic agent concentration with clinically acceptable accuracy within the first 5 min of the start of administration. The sequences are easily calculated in an Excel file and simple to use (one fixed FGF - FD setting), and will minimize agent consumption and reduce pollution by allowing to determine the lowest possible FGF that can be used. Different anesthesia machines will likely have different equations for different agents.

Highlights

  • The wide range of fresh gas flow - vaporizer setting (FGF - FD) combinations used by different anesthesiologists during the wash-in period of inhaled anesthetics indicates that the selection of FGF and FD is based on habit and personal experience

  • What fresh gas flow - vaporizer setting (FGF - FD) combination should be used for a particular patient when starting the administration of potent inhaled anesthetics to reach a target end-expired concentration (FA) after a predetermined time interval without excessively wasting potent inhaled anesthetic? The wide range of FGF - FD combinations used by different anesthesiologists during the wash-in period of potent inhaled anesthetics indicates that the selection of FGF and FD is based on habit and

  • Instead of relying on personal preference, we hypothesize that very specific FGF - FD combinations can be used in the individual patient to attain a FAt within a specified time interval by using an empirical model of the kinetics of inhaled anesthetics during wash-in

Read more

Summary

Introduction

The wide range of fresh gas flow - vaporizer setting (FGF - FD) combinations used by different anesthesiologists during the wash-in period of inhaled anesthetics indicates that the selection of FGF and FD is based on habit and personal experience. What fresh gas flow - vaporizer setting (FGF - FD) combination should be used for a particular patient when starting the administration of potent inhaled anesthetics to reach a target end-expired concentration (FA) after a predetermined time interval without excessively wasting potent inhaled anesthetic? We developed an empirical model for desflurane administered in O2 with an ADU - AS/5® anesthesia machine (Anesthesia Delivery Unit, General Electric, Helsinki, Finland) during the first 5 min of the anesthetic, and prospectively tested whether it allows the anesthesiologist to select a FGF - FD combination to attain a FAt of the inhaled anesthetic within a specified time interval (1 - 5 min) after turning on the vaporizer

Methods
Results
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.