Abstract

Aseptic loosening of the tibial component; which may be caused by mechanical stress shielding in the bone and may require revision surgery; is the primary concern of total knee replacement (TKR). The stiffness of the implant material had a marked influence on the stresses developed in the constituents and surrounding bones of the artificial knee and then will affect the bone stress shielding. Therefore, the functionally graded materials had been developed as a potential tibia tray material of TKR due to its improved capability of stress distribution. In the current investigation two dimensional finite element models have been developed to study bone and interface stresses for six different tibial prothesises (titanium, CoCrMo and four functional graded materials “FGM” models). The utilization of FGM tibia tray with elastic modulus changing gradually in vertical direction downwardly showed a favorable stress distribution outcome. Furthermore, the results has revealed that the FGM tibia tray will reduce the stress shielding in the surrounding bones of the artificial knee which will increase the life of the total knee prosthesis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.