Abstract
Ultra-high pressure liquid chromatography enables increased separation speed and efficiency. The quantitative improvement in efficiency is lower than that predicted by theory, and the reasons are not known. In this work, slow mass transport due to analyte desorption from the stationary phase is discussed as a possible contribution to the lower than expected efficiency. Data in the literature for the reversed phase elution of acetophenone, for which particle size was varied with constant particle composition and mobile phase, were used to test this possibility. The mass transport terms for the three particles sizes (1.7, 3.5 and 5.0 μm) fit well to a model that includes desorption from the stationary phase as a contribution, and this analysis yields an apparent desorption time constant of 2.0(±0.2) ms for acetophenone in a reversed-phase separation. The results indicate that it is reasonable to consider slow desorption as a possible contribution to the reduced plate height for sub-2-μm particles.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.