Abstract

Understanding the mass transfer behaviors in hollow fiber membrane module of artificial liver is important for improving toxin removal efficiency. A three-dimensional numerical model was established to study the mass transfer of small molecule bilirubin and macromolecule bovine serum albumin (BSA) in the hollow fiber membrane module. Effects of tube-side flow rate, shell-side flow rate, and hollow fiber length on the mass transfer of bilirubin and BSA were discussed. The simulation results showed that the clearance of bilirubin was significantly affected by both convective and diffusive solute transport, while the clearance of macromolecule BSA was dominated by convective solute transport. The clearance rates of bilirubin and BSA increasd with the increase of tube-side flow rate and hollow fiber length. With the increase of shell-side flow rate, the clearance rate of bilirubin first rose rapidly, then slowly rose to an asymptotic value, while the clearance rate of BSA gradually decreased. The results can provide help for designing structures of hollow fiber membrane module and operation parameters of clinical treatment.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.