Abstract

CO + and N 2 + are separated with resolution of 11,000 [full width half maximum (FWHM)] using a conventional quadrupole mass spectrometer by applying square wave voltages to the entrance and exit lenses to trap or reflect the ions for multiple passes. A resolution of 22,000 (FWHM) with 63% of the total signal remaining is attained using multiple passes when ions are stored between injection pulses. Gated ion extraction also reduces the mass shift and number and intensity of artifact peaks and permits better resolution compared to the performance obtained when the ions are injected continuously.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.