Abstract

The middle ear as a levered vibrating system for sound transmission from the external to the inner ear is affected by changes in ossicular chain mass. Mass loading of the ossicles may impair ossicular dynamics and sound transmission to the inner ear. It is incumbent on otologic surgeons and researchers of middle ear mechanics to consider the mass loading effect on middle ear function in clinical and physiological applications. The residual hearing and frequency response can change after surgery or implantation of middle ear prostheses. We conducted experiments on mass loading effects on the middle ear transfer functions by using laser Doppler interferometry and a human temporal bone model. Two implant mass loading conditions were tested on 17 fresh or fresh-frozen temporal bones and compared with the unloaded condition for the frequencies 250 to 8,000 Hz. The results show that the linearity of the middle ear function did not change, although displacement of the stapes footplate decreased after the increased masses were placed on the incudostapedial joint. The greater the mass of the implant, the less displacement was measured at the stapes footplate. We conclude that there is a quantitative limit to increased mass on the ossicular chain above which the mass will remarkably impair hearing thresholds.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.