Abstract

If an infant fails to initiate spontaneous breathing after birth, international guidelines recommend a positive pressure ventilation (PPV). However, PPV by face mask is frequently inadequate because of leak between the face and mask. Despite a variety of available face masks, none have been prospectively compared in a randomized fashion. We aimed to evaluate and compare leak between two commercially available round face masks (Fisher & Paykel (F&P) and Laerdal) in preterm infants <33 weeks gestational age in the delivery room. Infants born at the Royal Alexandra Hospital from April to September 2013 at <33 weeks gestational age who received mask PPV in the delivery room routinely had a flow sensor placed between the mask and T-piece resuscitator. Infants were randomly assigned to receive PPV with either a F&P or Laerdal face mask. All resuscitators were trained in the use of both face masks. We compared mask leak, airway pressures, tidal volume and ventilation rate between the two groups. Fifty-six preterm infants (n=28 in each group) were enrolled; mean±s.d. gestational age 28±3 weeks; birth weight 1210±448 g; and 30 (52%) were male. Apgar scores at 1 and 5 min were 5±3 and 7±2, respectively. Infants randomized to the F&P face mask and Laerdal face mask had similar mask leak (30 (25-38) versus 35 (24-46)%, median (interquartile range), respectively, P=0.40) and tidal volume (7.1 (4.9-8.9) versus 6.6 (5.2-8.9) ml kg(-1), P=0.69) during PPV. There were no significant differences in ventilation rate, inflation time or airway pressures between groups. The use of either face mask during PPV in the delivery room yields similar mask leak in preterm infants <33 weeks gestational age.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call