Abstract

The failure of facial prostheses is caused by limitations in the properties of existing materials, especially flexibility and durability. Therefore, this study evaluated the marginal deterioration of a silicone used for fabrication of facial prostheses (Silastic MDX4-4210, Dow Corning Corporation, Midland, MI, USA) according to the influence of artificial aging, daily disinfection, and 2 types of pigmentation. Thirty specimens were fabricated and subdivided in 6 groups: without pigmentation, pigmented with make-up powder and iron oxide, and evaluated with and without the action of the disinfectant. Analysis of marginal deterioration was performed on a scanning electron microscope (magnification, x5000) immediately 6 months and 1 year after fabrication of specimens, following the guidelines of ASTM International. After visual analysis of the photomicrographs, it was noticed that all groups presented marginal deterioration and alterations in surface texture with time. The use of disinfection did not contribute to the marginal deterioration of polymer (silicone), regardless of the pigmentation and artificial aging.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.