Abstract

Summary StatementUsing a simulated adult COVID-19 patient with hypoxemia, we investigated whether caregivers interrupting oxygen flow by manually occluding oxygen tubing with pliers during exhalation can conserve oxygen while maintaining oxygenation. Oxygen pinching reduced oxygen use by 51% to 64%, maintained simulated oxygen saturation between 88% and 90%, and increased simulated average alveolar partial pressure of oxygen from a room air baseline of approximately 131 to 294–424 mm Hg compared with 607 mm Hg with 10 liters per minute (LPM) continuous oxygen flow. Simulation provided a methodology to rapidly evaluate a technique that has begun to be used with COVID-19 patients in low-resource environments experiencing an acute oxygen shortage.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.