Abstract

The aim was to determine if dual-energy computed tomography (DECT) can discriminate ferromagnetic bullets from nonferromagnetic ones. The ferromagnetic properties of 29 bullets were determined. All bullets underwent a DECT evaluation, and a manual measurement of the dual-energy index was performed by 2 radiologists, followed by a fully automatic measurement, in 3 different positions in a phantom (air, water, and bone). Twenty-one bullets were found to be nonferromagnetic. There were significant differences between the mean of the dual-energy index measured in the core of ferromagnetic projectiles and those of nonferromagnetic ones in the lung position of the phantom (P = 0.007). Specificity was 86.36%, and negative predictive value was 90.48%. Among the 29 bullets, 24 were correctly classified. Our study demonstrates that DECT can assess the ferromagnetic status of bullets and their magnetic resonance imaging compatibility using an extended computed tomography scale in an ex vivo phantom model.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.