Abstract
The limits of accuracy for the quantitation of anti-D using the AutoAnalyzer have been reduced from +/- 26% with manually prepared dilutions to +/- 14% using an automatic pipette, as expressed by the 95% confidence limits. The error inherent in the AutoAnalyzer was estimated to contribute +/- 10% to the overall error. Problems associated with the reproducibility of this method for anti-D quantitation have been investigated, namely, the effect of the age of a given test cell, the use of different test cells, loss of sensitivity of the machine over a given day, and the reproducibility of results obtained at different positions on the standard graph.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.