Abstract

Tracheobronchial obstruction, although uncommon in the pediatric age group, remains a challenging problem. We review the long-term outcome of endoscopic metallic stenting in infants with tracheobronchial obstruction. Medical records of all pediatric surgical patients who underwent tracheobronchial metallic stenting in our center were reviewed retrospectively from 1996 to 2014. Patients' demographic data, including etiology, associated anomalies and nature of obstruction were reviewed. Outcome measures include complications such as re-stenosis, granulation tissue, stent migration, fractured stent, maximal tracheal diameter achieved, weaning of ventilator and growth at interval follow-up. Twelve balloon-expandable metallic stents were placed in the trachea (n = 10) and/or bronchi (n = 2) of 5 patients with a median age of 13 months (range 5-30 months). Etiology of the airway obstruction included congenital tracheal stenosis (n = 4), giant cervical and superior mediastinal lymphatic malformation with tracheobronchomalacia (n = 1). Seven complications were reported (3 patients developed granulation tissue, 2 patients had re-stenosis, 1 stent migrated, 1 stent fractured). All patients survived and were in good condition with a median follow-up of 16 years (range 11-18 years). Three patients weaned off ventilator and oxygen. Endoscopic stenting with metallic stent has satisfactory long-term outcome in treating infants with tracheobronchial obstruction.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call