Abstract

A rational systematic approach is presented in the evaluation of 14 men with persistent or recurrent urinary incontinence after placement of the AMS800 artificial urinary sphincter. Mechanical malfunctions, including device leaks and control assembly malfunctions, may often be discovered with physical examination and radiographic evaluations. Nonmechanical malfunctions, such as cuff erosion, inadequate cuff compression and functional urethral atrophy, can be diagnosed with perfusion sphincterometry, urethroscopy and a filling cystometrogram, all combined in 1 simplifed technique. With the presented algorithm the over-all success rate after 1 revision was 82%. Mechanical malfunctions appear to have a better revision rate of success than nonmechanical malfunctions (100% versus 71.5%, respectively).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.