Abstract

Failure of a balloon catheter to deflate is not uncommon and prevents its removal. Methods of overcoming the problem include traction, bursting the balloon by overinflation, dissolving it with solvents, puncturing it percutaneously with a needle, or puncturing it with a wire stylet passed through the catheter. All except the last technique have major disadvantages and are of questionable safety. Transcatheter puncture of the balloon was used in 16 patients to remove obstructed balloon catheters without any technical difficulty, distress to the patient, or complication. The procedure is safe, simple, and does not require an anaesthetic. If necessary it could be performed safely by nursing or paramedical staff without the patient having to be admitted to hospital. It is the method of choice for the management of this problem.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.