Abstract

Major obstetric haemorrhage (MOH) is a leading cause of maternal death and morbidity, with the majority of deaths occurring within four hours of delivery. Therefore, prompt identification of women at risk of MOH is crucial for the rapid assessment and management of blood loss to urgently restore haemodynamic stability. Furthermore, as the rate of blood loss during MOH can increase rapidly in the time when laboratory results are awaited, the management of MOH could benefit from point-of-care coagulation testing by the ROTEM analyser which has a quicker turnaround time compared to standard laboratory coagulation tests. A number of studies indicate that ROTEM-based management of MOH has resulted in a significant reduction in massive transfusions and decreased transfusion of concentrated red cells (CRC) and fresh frozen plasma (FFP) due to a reduction in total blood loss. Several reports which have linked MOH to the depletion of fibrinogen reserves indicate that the reduction in CRC and FFP transfusions is largely due to an increase in early fibrinogen replacement therapy which corrects hypofibrinogenemia. This short report discusses preliminary findings on the impact of ROTEM point-of-care haemostasis analyser on the transfusion of various blood products to obstetric women experiencing MOH at the Royal Gwent Hospital in South wales. The number of blood products transfused following decisions based on the ROTEM analyser measurements (ROTEM group) was compared to historical transfusion data before the ROTEM analyser became available (Pre-ROTEM group). Blood product transfusion in the Pre-ROTEM group was guided by measurements of standard laboratory coagulation tests in conjunction with the established major haemorrhage protocols at the time. The findings indicate that the ROTEM analyser was effective in managing MOH at point-of-care and led to a reduction in the transfusion of CRC, FFP and platelets. However, contrary to published studies, the reduction in blood product usage was not accompanied by an increase in fibrinogen replacement transfusion therapy, suggesting that the ROTEM’s FIBTEM assay accurately quantified fibrinogen levels based on fibrin-clot firmness to enable an early diagnosis of hypofibrinogenemia. Early establishment of the absence of hypofibrinogenemia helped to prevent unnecessary transfusion of fibrinogen concentrate in this study. These findings support the adoption of routine use of ROTEM analysers at point-of-care on labour wards to manage MOH and reduce fibrinogen replacement therapy. The ease of use and rapidity of ROTEM tests could enable departure from globally directed correction of coagulopathy during MOH to a more focussed and precise target transfusion therapy, which will ultimately reduce blood product wastage (including fibrinogen concentrate) whilst minimising transfusion-associated side effects such as alloimmunisation, circulatory overload and dilutional coagulopathy.

Highlights

  • Major obstetric haemorrhage (MOH) is a leading cause of maternal death and morbidity, with the majority of deaths occurring within four hours of delivery

  • The findings indicate that the ROTEM analyser was effective in managing MOH at point-of-care and led to a reduction in the transfusion of concentrated red cells (CRC), fresh frozen plasma (FFP) and platelets

  • Using the ROTEM analyser to regularly monitor and accurately determine the coagulation state of MOH patients before transfusion could be instrumental in informing goal-directed transfusion therapy, with potential to improve clinical outcomes through early fibrinogen replacement therapy which reduces the transfusion of high volumes of other blood products

Read more

Summary

Methods

ROTEM-guided usage of blood products for the management of MOH was evaluated at the Royal Gwent Hospital, in South Wales. Ethical approval was granted by the Aneurin Bevan University Health Board and Cardiff Metropolitan. ROTEM-based cases (ROTEM group) comprised of 108 women with MOH attending the main delivery unit (MDU) from January 2017 to May 2018. Non-ROTEM based cases (Pre-ROTEM group) comprised of 516 women with MOH who attended the MDU from January 2012 to December 2013. Pregnant women were included in the study if they experienced MOH (defined in this study as a blood loss

Results, discussion and conclusions
Findings
37. Johansson PI

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.