Abstract

Management of complications in long-term LVAD support like aortic regurgitation, progressive decline of right ventricular function, or pump thrombosis are challenging. After 2.5 years on a HeartMate II left ventricular assist device (LVAD) a 59-year-old patient was admitted to hospital due to severe hemolysis caused by a partial occlusion and thrombosis caused by a shift of the LVAD inflow cannula towards the lateral wall of the left ventricle (LV). Preoperative assessment revealed additional severe aortic regurgitation and impaired right ventricular (RV) function (severe tricuspid regurgitation, RV ejection fraction 25%, central venous pressure 14 mmHg). After LVAD device exchange and aortic valve replacement (AVR) RV failure occurred and a temporary right ventricular assist device (RVAD) was employed. RVAD explantation was feasible 21 days after surgery. The patient was discharged to a rehabilitation center. In this case, LVAD exchange with concomitant AVR and temporary RVAD efforts saved the patient's life and restored a good quality of life.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.