Abstract
This study investigates the incidence of full-field digital mammographic (FFDM) artifacts with three systems at two institutions and compares the artifacts between two detector types and two grid types. A total of 4,440 direct and 4,142 indirect FFDM images were reviewed by two radiologists, and artifacts were classified as patient related, hardware related, and software processing. The overall incidence of FFDM artifacts was 3.4% (292/8,582). Patient related artifacts (motion artifacts and skin line artifacts) were the most commonly detected types (1.7%). Underexposure among hardware related artifacts and high-density artifacts among software processing artifacts also were common (0.7 and 0.5%, respectively). These artifacts, specific to digital mammography, were more common with the direct detector type and the crossed air grid type than with the indirect type and linear grid type (p < 0.05). The most common mammographic artifacts on FFDM were patient related, which might be controlled by the instruction of a patient and technologist. Underexposure and high-density artifacts were more common with direct detector and crossed air type of grid.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.