Abstract

A simple and rapid method using a microhotplate-based preconcentrator and an ion mobility spectrometer (IMS) is proposed for the detection of malathion in water. The preconcentrator is prepared by micro-electro-mechanical system (MEMS) process. Coated with Polydimethylsiloxane (PDMS), it has the advantages of solvent-less, low energy cost, self-heating and ease to combine with IMS. The operating conditions of the preconcentrator-IMS system, such as extraction time, extraction temperature, agitation speed and desorption temperature, were optimised. Using the preconcentrator, the sampling procedure can be simplified and the detection limit of the system can be decreased. A linear relationship between the IMS response and the concentration of the analyte solution was verified. The malathion detection limit based on 3 times the baseline noise is 0.43 µg L−1 and the total analysis time is less than 30 minutes.