Abstract

Gamma-ray spectrometry losses through pulse processing dead time and pile-up are best assayed with an external pulse technique. In this work, the virtual pulse generator technique as implemented commercially with the Westphal loss free counting (LFC) module is set up and tested with four high resolution gamma-ray spectrometers. Dual source calibration and decaying source techniques are used in the evaluation of the accuracy of the correction technique. Results demonstrate the reliability of the LFC with a standardized conventional pulse processing system. The accurate correction during high rate counting, including during rapid decay of short lived activities, has been the basis for highly precise determinations in reference materials studies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.