Abstract
Image quality and radiation exposures of a mammographic technique using direct radiographic magnification at 2 X with a microfocal spot x-ray tube and a fast, double screen-film system were compared to those of conventional contact mammography with a rare-earth screen and molybdenum target tube. The results indicate that the magnification technique yields improved detection of microcalcifications and comparable visualization of soft-tissue details, with a large reduction in radiation exposure. This technique has demonstrated the feasibility of carrying out high-quality mammography with an entrance dose of 1.35 X 10(-3) Gy (135 mrad) for the average breast.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.