Abstract

In national and international bodies, there is active discussion of appropriate safety regulations of levels of magnetic field strength in MRI. Present limits are usually expressed in terms of the switching rate dB/dt, but the validity of this is open to debate. Application of the fundamental law of electrostimulation is well-established, both on theoretical and experimental grounds. Application of this law, in combination with Maxwell's law, yields a very simple equation that we call the fundamental law of magnetostimulation. This law has the hyperbolic form of a strength-duration curve and allows an estimation of the lowest possible value of the magnetic flux density capable of stimulating nerves and muscles. Calculations prove that the threshold for heart excitation is much higher than those for nerve and muscle stimulations. Experimental results from us and other authors confirm the correctness of the derived laws for magnetostimulation. In light of these findings, proposed safety limits should be reconsidered.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.