Abstract

QT dispersion is a measure of heterogeneity in ventricular repolarization. Increased ECG QT dispersion is associated with life-threatening ventricular arrhythmias. We studied if magnetocardiographic (MCG) measures of QT dispersion can separate postmyocardial infarction patients with and without susceptibility to sustained VT. Manual dispersion measurements were compared to a newly adapted automatic QT interval analysis method. Ten patients with a history of sustained VT (VT group) and eight patients without ventricular arrhythmias (Controls) were studied after a remote myocardial infarction. Single-channel MCGs were recorded from 42 locations over the frontal chest area and the signals were averaged. QT dispersion was defined as maximum-minimum or standard deviation of measured QT intervals. VT group showed significantly more QT and JT dispersion than Controls. QTapex dispersions were 127 +/- 26 versus 83 +/- 21 ms (P = 0.004) and QTend dispersions 130 +/- 37 versus 82 +/- 37 ms (P = 0.013), respectively. Automatic method gave comparable values. Their relative differences were 9% for QTapex and 27% for QTend dispersion on average. In conclusion, increased MCG QT interval dispersion seems to be associated with a susceptibility to VT in postmyocardial infarction patients. MCG mapping with automated QT interval analysis may provide a user independent method to detect nonhomogeneity in ventricular repolarization.