Abstract
In order to detect blood volume changes, we proposed the new magnetic plethysmographic sensor using time-varying magnetic fields, and thereby, measuring the change of impedance of exciting coil. The change of coil impedance was proportional to the change of blood volume passing through a volume-of-interests, which inductively coupled with exciting coil. In order to verify the feasibility of the proposed method, photo-plethysmographic and magneto-plethysmographic (MPG) signals were recorded from the index and middle finger of the left hand simultaneously. Comparison of the proposed sensor signals with ultrasound Doppler blood flow velocity recordings shows the excellent correlation (r = 0.9355, p <; 0.01) and the RR intervals derived from electrocardiograph and MPG signals showed a very high degree of correlation (r = 0.9823, p <; 0.01).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.