Abstract

Magnetization transfer (MT) techniques have been proposed as a method of increasing contrast in MR images. To evaluate the feasibility of MT imaging of the abdomen at 0.1 T and to assess the clinical utility of this technique, the authors studied tissue contrast with a gradient-echo pulse sequence and an MT sequence in four normal volunteers, and in 17 patients with known primary or secondary neoplasms of the liver. The MT technique increased contrast between the liver and other tissues such as spleen, skeletal muscle and subcutaneous fat. The technique also produced increased contrast between hepatic tumors and normal liver parenchyma in gradient-echo images.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.