Abstract

Advancements in brain–machine interfaces and neurological treatments urgently require the development of improved brain electrodes applied for long-term implantation, where traditional and polymer options face challenges like size, tissue damage, and signal quality. Carbon nanotubes are emerging as a promising alternative, combining excellent electronic properties and biocompatibility, which ensure better neuron coupling and stable signal acquisition. In this study, a new flexible brain electrode array based on 99.99% purity of single-walled carbon nanotubes (SWCNTs) was developed, which has 30 um × 40 um size, about 5.1 kΩ impedance, and 14.01 dB signal-to-noise ratio (SNR). The long-term implantation experiment in vivo in mice shows the proposed brain electrode can maintain stable LFP signal acquisition over 12 weeks while still achieving an SNR of 3.52 dB. The histological analysis results show that SWCNT-based brain electrodes induced minimal tissue damage and showed significantly reduced glial cell responses compared to platinum wire electrodes. Long-term stability comes from SWCNT’s biocompatibility and chemical inertness, the electrode’s flexible and fine structure. Furthermore, the new brain electrode array can function effectively during 7-Tesla magnetic resonance imaging, enabling the collection of local field potential and even epileptic discharges during the magnetic scan. This study provides a comprehensive study of carbon nanotubes as invasive brain electrodes, providing a new path to address the challenge of long-term brain electrode implantation.